forked from torvalds/linux
-
Notifications
You must be signed in to change notification settings - Fork 2
Commit
This commit does not belong to any branch on this repository, and may belong to a fork outside of the repository.
iio: documentation: Document scd4x calibration use
Add entries from Documentation/ABI/testing/sysfs-bus-iio-scd30 to Documentation/ABI/testing/sysfs-bus-iio. The attributes of the scd4x and scd30 are common. Remove Documentation/ABI/testing/sysfs-bus-iio-scd30. Signed-off-by: Roan van Dijk <[email protected]> Link: https://lore.kernel.org/r/[email protected] Signed-off-by: Jonathan Cameron <[email protected]>
- Loading branch information
Showing
2 changed files
with
41 additions
and
34 deletions.
There are no files selected for viewing
This file contains bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Original file line number | Diff line number | Diff line change |
---|---|---|
|
@@ -1958,3 +1958,44 @@ Description: | |
Specify the percent for light sensor relative to the channel | ||
absolute value that a data field should change before an event | ||
is generated. Units are a percentage of the prior reading. | ||
|
||
What: /sys/bus/iio/devices/iio:deviceX/calibration_auto_enable | ||
Date: June 2020 | ||
KernelVersion: 5.8 | ||
Contact: [email protected] | ||
Description: | ||
Some sensors have the ability to apply auto calibration at | ||
runtime. For example, it may be necessary to compensate for | ||
contaminant build-up in a measurement chamber or optical | ||
element deterioration that would otherwise lead to sensor drift. | ||
|
||
Writing 1 or 0 to this attribute will respectively activate or | ||
deactivate this auto calibration function. | ||
|
||
Upon reading, the current status is returned. | ||
|
||
What: /sys/bus/iio/devices/iio:deviceX/calibration_forced_value | ||
Date: June 2020 | ||
KernelVersion: 5.8 | ||
Contact: [email protected] | ||
Description: | ||
Some sensors have the ability to apply a manual calibration using | ||
a known measurement value, perhaps obtained from an external | ||
reference device. | ||
|
||
Writing a value to this function will force such a calibration | ||
change. For the scd30 the value should be from the range | ||
[400 1 2000]. | ||
|
||
Note for the scd30 that a valid value may only be obtained once | ||
it is has been written. Until then any read back of this value | ||
should be ignored. As for the scd4x an error will be returned | ||
immediately if the manual calibration has failed. | ||
|
||
What: /sys/bus/iio/devices/iio:deviceX/calibration_forced_value_available | ||
KernelVersion: 5.15 | ||
Contact: [email protected] | ||
Description: | ||
Available range for the forced calibration value, expressed as: | ||
|
||
- a range specified as "[min step max]" |
This file was deleted.
Oops, something went wrong.